Operation of the 586 generator

Procedure for turning on generator

  1. If not already turned on, the water used for cooling the generator needs to be turned on.

  2. Depress the "MAINS ON" button, and wait at least one (1) minute to allow the HV regulator tube to warm up and stabilize.

  3. Set the mA and kV selectors to minimum values if not already set to those values.

  4. Depress the "HT ON" button, keep the button depressed until all diagnostic LED's extinguish.

  5. Verify that the guages indicate thier set values and are stable.

  6. Select tube loading. For a Mo tube use 50kV and 20mA. Turn the KV's up first with mA's following. Take about 30-60 seconds to turn up each dial.

  7. Follow the procedure outlined in the Safety Plan marked as Single Crystal Diffractometer Operating Procedure.

Turning off the 586 generator.

  1. Slowly reduce mA setting to it's minimum value.

  2. Slowly reduce the kV setting to it's minimum value.

  3. Wait for some time for anode to cool down. A minute should suffice.

  4. Press "HT OFF" button.

  5. Wait for at least 8 seconds.

  6. Press "MAINS OFF".

Emergency Shutdown

To shutdown the generator in an emergency situation, press the MAINS OFF button. To avoid a collision of the circles turn off the MAINS button on the interface which controls the CAD4.